EMC testing cable coupling clamp jig - York EMC Services Ltd

Tel. +44 (0) 1904 324440

CCC01 Cable Clamp for EMC Testing

Home » Test Equipment » Cable Coupling Clamp

Cable Coupling Clamp 01

The Cable Coupling Clamp (CCC01) is a reusable test jig for the purpose of measuring the shielding properties of cables. The CCC01 is designed to allow easy positioning of the test and injection feed cables in accordance with the layout described in IEC 96-1 Amendment 2 1993-06. This simplifies the process of making repeatable measurements aimed at assessing the coupling and shielding effectiveness properties of a wide range of cables.

The CCC01 design is based on the details of the “launcher” arrangement, described in IEC 96-1 Amendment 2 1993-06 section A.5.6 Line Injection Method (frequency domain). This arrangement is also used in IEC 62153-4-6: 2006 and mandated for the line injection method in EN 50289-1-6.

 

Cable Coupling Clamp 01 Find your nearest distributor Request Quote

Features

   
Fixed cable routes •  Repeatable layout of test and injection cables 
Cable injection assemblies Injection cable runs are supplied preassembled and fitted with 
    ferrite common mode chokes
Supports a range of test cable sizes 2.5 mm, 5 mm and 10 mm aperture supports provided as 
    standard. Cable dimensions up to 22 mm x 13 mm are possible
    with the user-machinable fitting accessories
0.3 m & 0.5 m test lengths Base plate allows the injection assemblies to be mounted either 
    for a 0.3m cable length (EN 50289-1-6) or the extended 0.5 m 
    length  (IEC 96-1) for increased coupling distance
Configurable injection conductor Self-adhesive copper tape, easily trimmed to the correct width 
    required according to the circumference of the cable under test
Easy connection to test system Injection cables are terminated in N-type connectors

Applications

  • Cable shielding effectiveness measurements as indicated by:
    • IEC 96-1 Amendment 2 1993-06
    • IEC 62153-4-6:2006
    • EN 50289-1-6
  • Investigation of cable coupling phenomena

Specifications

Frequency Range •  10 kHz to 1 GHz typical (usable to higher frequencies, depending on the cables
     being tested and application of the test)
Output connectors •  50 Ω N-type sockets
Cable sizes •  Approx. 2.5 mm to 10 mm diameter using the three sets of fittings supplied.
  User-machinable fittings available as accessories, supporting maximum of cable
    dimension of 22 mm x 13 mm (rectangular cross-section) or a 13 mm diameter
Dimensions 220 mm × 880 mm (0.5 m test setting)
  220mm × 740mm (0.3m test setting)
Weight 3 kg (typical excluding cable-under-test and any associated ferrites)

 

Standard Kits

Part Number

Description

Parts included

CCC01KIT01 Standard CCC01 cable   •  Pair of preassembled cable injection assemblies
  coupling clamp kit •  Base plate with mounting positions for 0.3m and 
      0.5 m cable test lengths
       2.5 mm, 5 mm and 10 mm diameter cable mounting fittings
        Additional clip-on ferrite chokes
        Self-adhesive copper tape (signal injection conductor)
        Hard case for injection assemblies and accessories
    Manual
 
Please note: we are unable to supply the cable under test (CUT), any parts associated with its 
construction and any ferrite rings to be fitted to it.

Accessories:

CMF01 Kit of user-machinable blank fittings for custom cable apertures (single cable).
Special Contact us for information regarding our testing service for cables to both IEC 96-1 
  and EN 50289-1-6, using the CCC01.

 

Example Output Measurement Results